PM500 SERIES

EN60601-1 IEC60601-1 UL/cUL CB RoHS EAC cTUVus 62368-1 TÜV EN 62368-1

Feature

  • BF Class insulation
  • Operation altitude up to 5000 meters
  • 100-240 VAC input with active PFC
  • Less than 300 μA leakage current
  • EN55011 /55022 Class B conducted emissions
  • inhibit - TTL high to disable output
  • Compliant with RoHS requirements
  • Power consumption in standby mode less than 1W at standby power 5 V /100 mA

 

Model

Output

Efficiency

(typical)

115 /230 Vac

V1

Min.

Current

Max. Current

at convection

Max. Current

at 30 CFM

Tol.

Ripple & Noise

Max. Output

Power

PM500-12B

12 V

0 A

29.17 A

37.50 A

±2%

120 mV

350 W /450 W

88 /90%

PM500-13B

15 V

0 A

23.34 A

30.00 A

±2%

150 mV

350 W /450 W

88 /90%

PM500-13-1B

18 V

0 A

22.23 A

27.78 A

±2%

180 mV

400 W /500 W

88 /90%

PM500-14B

24 V

0 A

16.67 A

20.84 A

±2%

240 mV

400 W /500 W

89 /91%

PM500-15B

28 V

0 A

14.29 A

17.86 A

±2%

280 mV

400 W /500 W

89 /91%

PM500-17B

36 V

0 A

11.12 A

13.89 A

±2%

360 mV

400 W /500 W

89 /91%

PM500-18B

48 V

0 A

8.34 A

10.42 A

±2%

480 mV

400 W /500 W

89 /91%

PM500-19B

57 V

0 A

7.02 A

8.78 A

±2%

570 mV

400 W /500 W

89 /91%

Introduction

The PM500 series of AC-DC switching power supplies in a package of 4 x 7 x 1.7 inches are capable of delivering 450-500 watts of continuous power at 30 CFM forced air cooling or 350-400 watts at convection cooling. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover and fan assembly can be added during manufacturing. They are designed for medical applications including those needing BF rated insulation and/or an operation altitude up to 5000 meters.