Features
- Safety approval for Medical (IEC 60601-1) & ITE (IEC 62368-1)
- BF class insulation
- Operation altitude up to 5000 meters
- 2 x 4 inch footprint with 1.29 inch low profile
- Less than 175 μA leakage current
- Wide input range 80-264 VAC
- Meet EN 55011 / 55032 Class B
- Short-circuit protection
- No load power consumption less than 0.15W
- Compliant with RoHS requirement
|
|
Output |
Efficiency (typical) 115/230 Vac |
||||||
|
Model |
V1 |
Min. load |
Max. Current at convection |
Max. Current at 7.5 CFM |
Tol. |
Ripple & Noise |
Max. Power |
|
|
PM101-12A |
12 V |
0 A |
6.67 A |
8.34 A |
±2% |
120 mV |
80 W /100 W |
87 /90% |
|
PM101-13A |
15 V |
0 A |
5.34 A |
6.67 A |
±2% |
150 mV |
80 W /100 W |
87 /90% |
|
PM101-13-1A |
18 V |
0 A |
4.45 A |
5.56 A |
±2% |
180 mV |
80 W /100 W |
87 /90% |
|
PM101-14A |
24 V |
0 A |
3.34 A |
4.17 A |
±2% |
240 mV |
80 W /100 W |
88 /90% |
|
PM101-15A |
28 V |
0 A |
2.86 A |
3.58 A |
±2% |
280 mV |
80 W /100 W |
88 /90% |
|
PM101-16-1A |
32 V |
0 A |
2.50 A |
3.13 A |
±2% |
320 mV |
80 W /100 W |
88 /90% |
|
PM101-17A |
36 V |
0 A |
2.23 A |
2.78 A |
±2% |
360 mV |
80 W /100 W |
88 /90% |
|
PM101-18A |
48 V |
0 A |
1.67A |
2.09 A |
±2% |
480 mV |
80 W /100 W |
88 /90% |
Introduction