PM150 SERIES

EN60601-1 IEC60601-1 UL/cUL CB RoHS EAC

Features

  • BF Class insulation
  • No load power consumption less than 0.5W without PFD or 1W with PFD
  • Less than 275 μA leakage current
  • Operation altitude up to 5000 meters
  • 2 x 4 inch footprint with 1.3 inch low profile
  • Wide input range 80-264 VAC
  • Meet EN55011 /55022 and FCC Class B
  • Power Factor 0.98 typical
  • 100% burn-in at full load 
  • Short-circuit & Over-temperature protection
  • Power Fail Detect (PFD) signal (optional) 

 

 

Output

  Model

V1

Min. load

Max. Current

at convection

Max. Current

at 7.5 CFM

Peak

Current

Tol.

Ripple &

Noise

Max. Power

PM150-12A

12 V

0 A

8.35 A

12.50 A

14.0 A

±2%

120 mV

100 W /150 W

PM150-13A

15 V

0 A

6.70 A

10.00 A

11.0 A

±2%

150 mV

100 W /150 W

PM150-13-1A

18 V

0 A

5.56 A

 8.34 A

 9.2 A

±2%

180 mV

100 W /150 W

PM150-14A

24 V

0 A

4.20 A

 6.25 A

 7.0 A

±2%

240 mV

100 W /150 W

PM150-16A

30 V

0 A

3.34 A

 5.00 A

 5.6 A

±2%

300 mV

100 W /150 W

PM150-17A

36 V

0 A

2.78 A

 4.17 A

 4.6 A

±2%

360 mV

100 W /150 W

PM150-18A

48 V

0 A

2.10 A

 3.13 A

 3.5 A

±2%

480 mV

100 W /150 W

Introduction

PM150 series of AC-DC switching power supplies in a package of 2 x 4 x 1.3 inches are capable of delivering 100-150 watts of continuous power at 7.5 CFM forced air cooling or 100 watts at convection cooling. The units are constructed on a printed circuit board. They are specially designed for medical applications. The units are certified also to IEC /EN /UL /CSA 60950-1 and suitable for data networking, industrial and telecommunication applications.